Ventilator

Product Name:

Ventilator

Detailed Introduction:

portable ventilator / icu ventilator / oxygen ventilator

Positive Airway Pressure Device

Ventilator.jpg

.Mobile Bi-level for hospital

ventilator11.jpg

ventilator12.jpg

1. Maximum inhale pressure 30cmH2O, with separate humidifier, better humidification
2. With ST and target tidal volume mode settings
3. Fast rise time
4. Normal inhalation intensity, respiration rate is 54 times / min, and each exhalation triggers the ventilator perfectly. Easily keep in sync
5. Good synchronization with big flow

30.jpg

Technical parameters of mobile bi-level PAP device
• 1. Scope of application:
For adult patients with ARDS, hypoventilation syndrome, for bi-level non-invasive ventilation in a hospital or home.
• 2. the main function:
1. Different treatment options can be provided according to patients’ needs, from CPAP to S, T, S / T, VGPS five modes。
2. Adjustable trigger sensitivity.
3. Real-time monitoring and recording of blood oxygen saturation.
4. alarm function.
5. Record function (SD card).
6. Maximum leak compensation function. 60L/min
7. Waveform display in screen :Time&pressure, time&flow
8. Medical use separate humidifier, better humidification and heating
9. with Trolley, easy to move the device.
• 3. the key technical parameters:
1. 1. Working mode: CPAP, S, T, ST, VGPS.
2. 2. Ventilation parameters:
• 2.1 Inspiratory pressure: 4-30 cmH2O (0.5 cmH2O interval)
• 2.2 Expiratory pressure: 4-30 cmH2O (0.5 cmH2O interval)
• 2.3 Treatment pressure: 4-20 cmH2O (in CPAP mode, 0.5 cmH2O interval)
• 2.4 Delayed boost: 0-45 minutes (interval of 5 minutes)
• 2.5 Breathing frequency: 5-50 times / minute (interval 1 time / minute)
• 2.6 Breathing ratio: 10-70% (interval of 1%)
• 2.7 Rise time: 1-5 steps (1 step interval, 1 step rises fastest, 5 steps rises most slowly)
• 2.8 Inhalation sensitivity: 1-5 steps (1 step interval, 1st step triggers fastest, 5th step 1st step triggers slowest)
• 2.9 Exhalation sensitivity: 1-5 (1 interval, the 1st trigger is the fastest, 5th and 1st the slowest)
• 2.10 Tidal volume: 50-1500ml (effective in VGPS mode, interval 50ml)
• 3. VGPS(same with AVAPS,can set Target tidal volume)mode:
• 3.1 Maximum IPAP value adjustable range: 4-30 cmH2O
• 3.2 Adjustable range of minimum IPAP value: 4-30 cmH2O
• 3.3 EPAP value adjustable range: 4-30 cmH2O
• 3.4 Target Tidal Volume: 50-1500ml
• 4. Display range of monitoring setting parameters:
• 4.1 Respiratory waveform can be displayed;
• 4.2 Current mode: CPAP, S, T, ST, VGPS
• 4.3 Inspiratory pressure: 4-30 cmH2O
• 4.4 Expiratory pressure: 4-30 cmH2O
• 4.5 Breathing frequency: 5-50 times / minute
• 4.6 Breathing ratio: 10-70%
• 4.7 Air leakage: 20-99.9lpm
• 4.8 Tidal volume: 50-1500ml
• 4.9 minute ventilation: 0-50lpm
• 4.10 Blood oxygen saturation: 70-100%
• 4.11 Pulse rate: 25-250bpm
• 4.12 Delay boost: 0-45 minutes
• 5. Alarm items:
• 5.1 High pressure sound prompt: Turn on / off sound prompt
• 5.2 Low minute ventilation sound prompt: Turn on / off sound prompt, can be set to 1-10L (interval 1L)
• 5.3 Air leakage sound prompt: on / off
• 5.4 Power off sound prompt: on / off
• 5.5 Choking voice prompt: On / off, choking time can be set to 10-40 seconds (interval of 5 seconds)
• 5.6 Low Tidal Volume Sound Prompt: On / Off, Set Tidal Volume Value Range 50-500ml ((Interval 50ml))
• 6. Specifications
• 6.1 Weight: about 1500g
• 6.2 Operating temperature: 5-35 ° C
• 6.3 Relative humidity: 10% -90%
• 6.4 Power supply: input is AC 100-240V, 50-60Hz, 1.8A Max; output: DC 24V, 3.33A.