510(K)
SUMMARY
Submitter’s Name: HERS MEDICAL CO., LTD
7F-4 NO : 246 SEC .2 CHANG-AN
EAST RD.,
TAIPEI TAIWAN
Date summary
prepared:
Sept. 10, 2004
Device Name:
Propietary Name: HERS MEDICAL Blood Pressure Monitor, Model
No: YS-796, YS-818, YS-793,
YS-794, YS-795, YS-805, YS-807 and same series model number.
Common or Usual Name: NONINVASIVE BLOOD PRESSURE MEASUREMENT
SYSTEM
Classification Name: Blood Pressure Monitor, Class II,
CFR 870.1130
Indications for Use:
The above mentioned device is a noninvasive blood pressure
measurement system intended to measure the systolic and diastolic blood
pressure and pulse rate of an adult individual, over age 18,
At home by using a non-invasive technique in which an
inflatable cuff is wrapped around the wrist
And Arm. The cuff circumference is limited to be 5.25–
7.75.
Description of the device:
HERS MEDICAL Blood Pressure Monitor, Model No: YS-796,
YS-818, YS-793, YS-794, YS-795,
YS-805,
YS-807 and same series model number.
Uses the Oscillometric method to measure the blood
pressure. The Oscillometric method is adopted clinically to measure the
blood pressure recently.
It is not needed to use the stechoscope, as in the traditional measuring
method, to monitor the
Korotkov sound when deciding the systolic or diastolic pressure. The
Oscillometric method senses
the vibrating signal via the closed air pipe system. And utilizes the
microcomputer to automatically
sense the characteristics of the pulse signal. Through simple
calculation, clinically proven, the
reading can reflect the accurate real blood pressure, and the systolic
pressure is defined as the
pressure when the cuff pressure Oscillating amplitude begins to increase
and the diastolic pressure as
the pressure when the cuff pressure oscillating amplitude stops
decreasing.
Performance Testing:
Electric Safety Requirement Test Report of EN 60601-1:1990 & EN
1060-1/EN1060-3,
And EMC test report of EN 60601-1-2(EN 55011:1991and EN 6100-4-2:1995)
ANSI/AAMI SP10-1992 Electronic or Automated Sphygmomanometers
Legally marketed device for substantial equivalence comparison:
EikonAutomatic Digital Blood Pressure Monitor, HD-400M(K021239)
Summary for substantial equivalence comparison:
Same characteristics: intended use, technological characteristics, power
supply, display, measuring
Range, accuracy, operating and storage environments.
Different characteristics: memory, dimensions, weight, and storage
environments
1.
As we can understand, the memory feature is to memorize the measurement
data taken
Previously and is related to the usage convenience, not to raise any
safety or effectiveness hazard.
2.
The differences between dimensions and weight are related to the
designing aspects.
These differences are not to raise any safety or effectiveness aspect.
They are decided to be substantially equivalent.
of Pen Shape